Steam generator

Burns Pressure equipment/vessels Other risk levels
Photo of recalled product: Steam generator
Alert NumberOR/00067/25
BrandTRANS MEDIKAL - GETINGE
CategoryPressure equipment/vessels
Risk TypeBurns
Notifying CountryFrance
Country of OriginTürkiye
Model Number20 kW ; 22,5 kW ; 30 kW ; 37,5 kW ; 40 kW ; 45 kW ; 50 kW ; 60 kW
Published9 May 2025

Risk Description

Due to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns. The product does not comply with the requirements of the Pressure Equipment Directive.

Measures Taken

Type of economic operator to whom the measure(s) were ordered: DistributorCategory of measure(s): Recall of the product from end usersDate of entry into force: 02/03/2025Type of economic operator taking notified measure(s): DistributorCategory of measure(s): Making the marketing of the product subject to prior conditionsDate of entry into force: 02/03/2025

Product Description

Steam generator integrated into steriliser.

View on EU Safety Gate →