FFP2 respiratory mask high filtration efficiency

Health risk / other Protective equipment Serious risk
Photo of recalled product: FFP2 respiratory mask high filtration efficiency
Alert NumberA12/01531/21
BrandTEX-TECH
CategoryProtective equipment
Risk TypeHealth risk / other
Notifying CountryGermany
Country of OriginCzechia
Model NumberFFP2 Premium
Published12 November 2021

Risk Description

The particle/filter retention of the material is insufficient (measured values: as low as 83%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. Furthermore, the physical resistance of the filtering material is too high (measured values of exhalation resistance: up to 3,8 mbar). Consequently, it may prevent a sufficient quantity of air to be exhaled, causing difficulties to breathe. The product does not comply with the Personal Protective Equipment (PPE) Regulation and with the relevant European standard EN 149.

Measures Taken

Type of economic operator to whom the measure(s) were ordered: OtherCategory of measure(s): Ban on the marketing of the product and any accompanying measuresDate of entry into force: UnknownType of economic operator taking notified measure(s): RetailerCategory of measure(s): Recall of the product from end usersDate of entry into force: Unknown

Product Description

White masks with ‘TEX-TECH FFP2 Premium’ labelling. The number 1040 appears next to the CE mark on the masks.

View on EU Safety Gate →