BTL Respirator FLAT-FIT Healthcare Respirator BTL

Health risk / other Protective equipment Serious risk
Photo of recalled product: BTL Respirator FLAT-FIT Healthcare Respirator BTL
Alert NumberA12/00431/21
BrandBTL Respirator
CategoryProtective equipment
Risk TypeHealth risk / other
Notifying CountryGermany
Country of OriginUnknown
Model NumberFLAT FIT
Published2 April 2021

Risk Description

The particle/filter retention of the material is insufficient (measured values: as low as 82,5 %).Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. The product does not comply with the PPE Regulation and the relevant European standard EN 149.

Measures Taken

Type of economic operator to whom the measure(s) were ordered: OtherCategory of measure(s): Warning consumers of the risksDate of entry into force: 31/01/2021Type of economic operator taking notified measure(s): RetailerCategory of measure(s): Removal of this product listing by the online marketplaceDate of entry into force: 01/02/2021Type of economic operator taking notified measure(s): RetailerCategory of measure(s): Recall of the product from end usersDate of entry into force: 01/02/2021Type of economic operator taking notified measure(s): ManufacturerCategory of measure(s): Withdrawal of the product from the marketDate of entry into force: 11/02/2021

Product Description

White particle filter mask. The product was also sold online.

View on EU Safety Gate →